332
Maintenance and troubleshooting
Type 8098 FLOWave L
24.10.3
Message "Calibration cancelled"
Message
Calibration cancelled
Symbol displayed in the
information bar
Possible cause
The user has interrupted a calibration of the zero flow, before the waiting
time of 30 seconds has elapsed.
What to do?
→
Do a new calibration and observe the given instructions.
24.10.4
Message "Calibration cancelled, the flow rate is less than 5 %
of the full scale"
Message
Calibration cancelled, the flow rate is less than 5 % of the full scale
Symbol displayed in the
information bar
Possible cause
The calibration has failed because
the flow rate is less than 5 % of the full
scale.
What to do?
→
Make sure the flow rate is higher than 5 % of the full scale.
→
Do a new calibration.
24.10.5
Message "Resulting K factor is less than 0.8 or higher than
1.2"
Message
Resulting K factor is less than 0.8 or higher than 1.2
Symbol displayed in the
information bar
Possible cause
When calibrating the K factor by using a teach-in procedure depending
on the flow rate or depending on a known volume, you have entered a ref
-
erence value that varies for ±20 % from the measured value.
What to do?
→
Do a new calibration.
→
Enter a correct reference value.
24.10.6
Message "Resulting offset is higher than 10 °C, 18 °F"
Message
Resulting offset is higher than 10 °C, 18 °F
Symbol displayed in the
information bar
Possible cause
When calibrating the offset value of the liquid temperature, you have entered
a reference value that varies for ±10 °C (18 °F) from the measured value.
What to do?
→
Do a new calibration.
→
Enter a correct reference value.
English
Summary of Contents for Flowave 8098
Page 4: ...General contents 4 Type 8098 FLOWave L English...
Page 12: ...12 Type 8098 FLOWave L English...
Page 36: ...36 Technical data Type 8098 FLOWave L English...
Page 72: ...72 Type 8098 FLOWave L English...
Page 78: ...78 Commissioning Type 8098 FLOWave L English...
Page 96: ...96 Type 8098 FLOWave L English...
Page 144: ...144 Type 8098 FLOWave L English...
Page 248: ...248 Type 8098 FLOWave L English...
Page 286: ...286 SAW sensor Maintenance Type 8098 FLOWave L English...
Page 304: ...304 Type 8098 FLOWave L English...
Page 316: ...316 Type 8098 FLOWave L English...
Page 350: ...350 Spare parts and accessories Type 8098 FLOWave L English...
Page 352: ...352 Appendix Type 8098 FLOWave L English...
Page 353: ......
Page 354: ...www burkert com...