329
Maintenance and troubleshooting
Type 8098 FLOWave L
24.9.15
Message "Measure board is in boot starter mode, no firmware
found n°1"
Message
Measure board is in boot starter mode, no firmware found n°1
Symbol displayed in the
information bar
Possible cause
The firmware of the measurement board is lost or is not valid.
What to do?
→
Start the device again.
→
If the error is still there, send the device back to Bürkert.
24.9.16
Message "Measured values cannot be used"
Message
Measured values cannot be used
Symbol displayed in the
information bar
-
Possible cause
The device cannot measure the liquid parameters, for example, because
there are too many bubbles in the liquid or the sensor tube is not completely
filled.
What to do?
→
Make sure there is no problem in the installation.
24.9.17
Message "Communication between transmitter PCB and
measurement PCB has been interrupted n°x"
Message
Communication between transmitter PCB and measurement PCB has been
interrupted n°x
Symbol displayed in the
information bar
Possible cause
There is no communication between the sensor and the transmitter.
What to do?
→
Make sure the cable connecting the sensor to the transmitter is not
broken and correctly plugged in. See chapter 7.3.1 Changing the
position of the transmitter on the sensor.
24.9.18
Message "The measurement board bootloader operation
failed n° 1"
Message
The measurement board bootloader operation failed n° 1
Symbol displayed in the
information bar
Possible cause
During the firmware update, the sensor software could not be updated.
What to do?
→
Send the device back to Bürkert.
English
Summary of Contents for Flowave 8098
Page 4: ...General contents 4 Type 8098 FLOWave L English...
Page 12: ...12 Type 8098 FLOWave L English...
Page 36: ...36 Technical data Type 8098 FLOWave L English...
Page 72: ...72 Type 8098 FLOWave L English...
Page 78: ...78 Commissioning Type 8098 FLOWave L English...
Page 96: ...96 Type 8098 FLOWave L English...
Page 144: ...144 Type 8098 FLOWave L English...
Page 248: ...248 Type 8098 FLOWave L English...
Page 286: ...286 SAW sensor Maintenance Type 8098 FLOWave L English...
Page 304: ...304 Type 8098 FLOWave L English...
Page 316: ...316 Type 8098 FLOWave L English...
Page 350: ...350 Spare parts and accessories Type 8098 FLOWave L English...
Page 352: ...352 Appendix Type 8098 FLOWave L English...
Page 353: ......
Page 354: ...www burkert com...