Outputs – Parameter
Type 8098 FLOWave L
18.3
Setting the parameters of an analogue output
By default, the device has 2 analogue outputs,
Output 1:analog
and
Output 3:analog
. You can change the
type of the output 3: see chapter 18.2
The following parameters can be set:
• the
process value
associated to the analogue output.
•
the value of the process variable, which is associated to the 4 mA current of the analogue output.
•
the value of the process variable, which is associated to the 20 mA current of the analogue output.
• the
damping
level of the values that are transmitted on the analogue output. By default, the values trans-
mitted on the analogue output are not damped.
• the behaviour of the analogue output depending on the status of the device.
Table 42:
Default parameters of the 2 analogue outputs
Parameter
Default value
Process value
associated to the analogue output
Volume flow rate
4 mA value
0.0 l/min
20 mA value
Full scale of the flow rate measurement range. Value
depends on the DN of the process connections.
Damping
level
None
Behaviour if a
Failure
message is generated by the
device
22mA
Behaviour if an
Out of spec.
message is generated
by the device
Continue
Behaviour if a
Maintenance req.
message is
generated by the device
Continue
18.3.1
Changing the process value and the process value range
associated to an analogue output
To change the process value and the process value range associated to an analogue output, do the
following:
→
Go to the
CONFIGURATION
view.
→
Outputs
→
Confirm to access the
Parameter
view.
→
Output 1: analog
or
Output 3: analog
→
Settings
The
Current settings
are displayed
→
Choose a process value
→
Set the value associated to a 4 mA current
→
Set the value associated to a 20 mA current
The
New settings
are displayed.
→
Save.
The process value and the process value range associated to the analogue output are changed.
English
Summary of Contents for Flowave 8098
Page 4: ...General contents 4 Type 8098 FLOWave L English...
Page 12: ...12 Type 8098 FLOWave L English...
Page 36: ...36 Technical data Type 8098 FLOWave L English...
Page 72: ...72 Type 8098 FLOWave L English...
Page 78: ...78 Commissioning Type 8098 FLOWave L English...
Page 96: ...96 Type 8098 FLOWave L English...
Page 144: ...144 Type 8098 FLOWave L English...
Page 248: ...248 Type 8098 FLOWave L English...
Page 286: ...286 SAW sensor Maintenance Type 8098 FLOWave L English...
Page 304: ...304 Type 8098 FLOWave L English...
Page 316: ...316 Type 8098 FLOWave L English...
Page 350: ...350 Spare parts and accessories Type 8098 FLOWave L English...
Page 352: ...352 Appendix Type 8098 FLOWave L English...
Page 353: ......
Page 354: ...www burkert com...