197
SAW sensor – Parameter
Type 8098 FLOWave L
15.9.4
Deactivating the damping of the liquid density values
By default, the measured liquid density values are damped with the level
Medium
.
To deactivate the damping of the liquid density values, do the following:
→
Go to the
CONFIGURATION
view.
→
SAW sensor
→
Confirm to access the
Parameter
view.
→
Stand. meas. values
→
Density
→
Damping
The
Current settings
are displayed
→
Choose
None
The
New settings
are displayed.
→
Save.
The damping of the liquid density values is inactive.
15.9.5
Activating the monitoring of the liquid density
Because of a malfunction in the process or in the flow rate sensor, the measured liquid density value can be
too high or too low.
A monitored value can be:
• in the normal operating range.
• in the warning range,
• in the error range.
You can set 4 limit values: 2 error limits and 2 warning limits.
→
To set the limit values, see chapter 15.9.7 Changing the error limits, the warning limits and the hysteresis
15.4.5 explains how the device reacts when the monitored value enters in another range (for
example, from the normal range into the warning range). The reaction time depends on the hysteresis value
and if the monitored value increases or decreases.
By default, the monitoring of the liquid density and the diagnostics are all disabled.
To activate the monitoring of the liquid density, do the following:
→
Go to the
CONFIGURATION
view.
→
SAW sensor
→
Confirm to access the
Parameter
view.
→
Stand. meas. values
→
Density
English
Summary of Contents for Flowave 8098
Page 4: ...General contents 4 Type 8098 FLOWave L English...
Page 12: ...12 Type 8098 FLOWave L English...
Page 36: ...36 Technical data Type 8098 FLOWave L English...
Page 72: ...72 Type 8098 FLOWave L English...
Page 78: ...78 Commissioning Type 8098 FLOWave L English...
Page 96: ...96 Type 8098 FLOWave L English...
Page 144: ...144 Type 8098 FLOWave L English...
Page 248: ...248 Type 8098 FLOWave L English...
Page 286: ...286 SAW sensor Maintenance Type 8098 FLOWave L English...
Page 304: ...304 Type 8098 FLOWave L English...
Page 316: ...316 Type 8098 FLOWave L English...
Page 350: ...350 Spare parts and accessories Type 8098 FLOWave L English...
Page 352: ...352 Appendix Type 8098 FLOWave L English...
Page 353: ......
Page 354: ...www burkert com...