343
Maintenance and troubleshooting
Type 8098 FLOWave L
24.12.6
Message "Low flow cut off"
Message
Low flow cut off
Symbol displayed in the
information bar
Depends on the device status the event is associated to.
Possible cause
The cut-off value of the flow rate has been used.
The cut-off function must be enabled. See chapter 15.4.9 Enabling the
cut-off function of the volume flow rate
.
The related event must be enabled. See chapter 15.14 Diagnostics: moni
toring special events that occur in the process, on the sensor or on the
What to do?
→
If necessary, increase the flow rate value until it is higher than the cut-off
value.
24.12.7
Message "Change of liquid"
Message
Change of liquid
Symbol displayed in the
information bar
Depends on the device status the event is associated to.
Possible cause
A different liquid flows in the pipe.
The message is active for 10 s on the display.
The related event must be enabled. See chapter 15.14 Diagnostics: moni
toring special events that occur in the process, on the sensor or on the
What to do?
→
Make sure the liquid flowing in the pipe is the correct one.
24.12.8
Message "Backward flow"
Message
Backward flow
Symbol displayed in the
information bar
Depends on the device status the event is associated to.
Possible cause
The liquid flows in the opposite direction as the one set in chapter 17.4
Setting the direction of the flow
.
The related event must be enabled. See chapter 15.14 Diagnostics: moni
toring special events that occur in the process, on the sensor or on the
What to do?
→
Make sure the liquid flows in the correct direction.
English
Summary of Contents for Flowave 8098
Page 4: ...General contents 4 Type 8098 FLOWave L English...
Page 12: ...12 Type 8098 FLOWave L English...
Page 36: ...36 Technical data Type 8098 FLOWave L English...
Page 72: ...72 Type 8098 FLOWave L English...
Page 78: ...78 Commissioning Type 8098 FLOWave L English...
Page 96: ...96 Type 8098 FLOWave L English...
Page 144: ...144 Type 8098 FLOWave L English...
Page 248: ...248 Type 8098 FLOWave L English...
Page 286: ...286 SAW sensor Maintenance Type 8098 FLOWave L English...
Page 304: ...304 Type 8098 FLOWave L English...
Page 316: ...316 Type 8098 FLOWave L English...
Page 350: ...350 Spare parts and accessories Type 8098 FLOWave L English...
Page 352: ...352 Appendix Type 8098 FLOWave L English...
Page 353: ......
Page 354: ...www burkert com...