5-80 Waterfall
Commands
SR785 Dynamic Signal Analyzer
Waterfall Commands
WDSP (?) d {, i}
The WDSP command sets (queries) the Waterfall Display Mode for display d.
The parameter i selects Normal (0) or Waterfall (1).
This command is not valid when the Measurement Group is Swept Sine.
WSTO (?) d {, i}
The WSTO command sets (queries) the Waterfall Storage for display d. The
parameter i selects Off (0) , Continuous (1), or One Shot (2). The set command
requires d-=2 (both displays).
This command is not valid when the Measurement Group is Swept Sine.
WFSB (?) d {, i}
The WFSB command sets (queries) the Waterfall Save Option for display d. The
parameter i selects Save All Measurements (0) or Save Active Measurement
Only (1).
This command is not valid when the Measurement Group is Swept Sine.
WTOT (?) d {, i}
The WTOT command sets (queries) the Waterfall Total Count for display d. The
parameter i is a number of records. If the allocated memory is too small for the
requested number of records, an error occurs.
This command is not valid when the Measurement Group is Swept Sine.
WAVA ? d
The WAVA command queries the number of records available in the waterfall
buffer for display d.
Records are not stored in the buffer unless Waterfall Storage is on (Continuous
or One Shot).
This command is not valid when the Measurement Group is Swept Sine.
WFSK (?) d {, i}
The WFSK command sets (queries) the Waterfall Skip count for FFT
measurements for display d. The parameter i is a number of records.
This command is valid only when the Measurement Group is FFT, Correlation,
Order or Time/Histogram.
WOSK (?) d {, x} <ks, [s], ms>
The WOSK command sets (queries) the Waterfall Skip count for Octave
measurements for display d. The parameter x is a time from .008 to 1000 s.