Frequency
Commands
5-33
SR785 Dynamic Signal Analyzer
This command is valid only when the Measurement Group is Swept Sine. The
set command requires a display to be Live.
SNPS (?) d {, i}
The SNPS command sets (queries) the swept sine Number Of Points of display
d. The parameter i is a number of points from 10 to 2047. The set command
requires d=2 (both displays).
This command is valid only when the Measurement Group is Swept Sine. The
set command requires a display to be Live.
SSKP (?) d {, i}
The SSKP command sets (queries) the swept sine Maximum Number Of Skips
of display d. The parameter i is a maximum number of skips from 2 to 200. The
set command requires d=2 (both displays).
This command is valid only when the Measurement Group is Swept Sine. The
set command requires a display to be Live.
SFST (?) d {, x}
The SFST command sets (queries) the swept sine Faster Threshold of display d.
The parameter x is a level from 0.01 to 3.0 (dB). The set command requires d=2
(both displays).
This command is valid only when the Measurement Group is Swept Sine. The
set command requires a display to be Live.
SSLO (?) d {, x}
The SSLO command sets (queries) the swept sine Slower Threshold of display d.
The parameter x is a level from 0.05 to 6.0 (dB). The set command requires d=2
(both displays).
This command is valid only when the Measurement Group is Swept Sine. The
set command requires a display to be Live.