Order Frequency Menu 4-21
SR785 Dynamic Signal Analyzer
Storage
Determines the storage mode for track measurements. If Storage is set to One Shot, the
track measurement will end after the specified number of points have been acquired. If
Storage is set to continuous, the measurement will not stop, but the new data will
overwrite the oldest data.
Command: OSTO (?) d {, i}
Track 1 Order
Specifies the order associated with the Track 1 measurement. If the entered value is not
an integer multiple of delta order, the SR785 will display the closest available order to
the entered value.
Regardless of the order specified, the SR785 always tracks all orders specified by the
max order and delta order softkeys. The Track 1 Order softkey merely selects which
order will displayed. Once a track measurement is done or paused, any order track can be
displayed without retaking any data.
Command: O1TK (?) d {, f}
Input Channel (Track 1)
Specifies the input channel associated with the Track 1 measurement.
Regardless of the input channel specified, the SR785 always tracks all orders specified
by the max order and delta order softkeys for both input channels. The Input Channel
(Track 1) softkey merely selects which channel will be displayed. Once a track
measurement is done or paused, data for each channel can be displayed without retaking
any data.
Command: O1BN (?) d {, i}
Track 2 Order
Specifies the order associated with the Track 2 measurement. If the entered value is not
an integer multiple of delta order, the SR785 will display the closest available order to
the entered value.
Regardless of the order specified, the SR785 always tracks all orders specified by the
max order and delta order softkeys. The Track 2 Order softkey merely selects which
order will displayed. Once a track measurement is done or paused, any order track can be
displayed without retaking any data.
Command: O2TK (?) d {, f}