3-36 Status
Indicators
SR785 Dynamic Signal Analyzer
In Octave group, [Span Down] moves the octave measurement span of both displays
down by an octave.
[Span Down] affects each display independently if their frequency parameters are
unlinked.
[Span Down] has no affect for Swept Sine measurements, Order measurements, or
Time/Histogram measurements.
[Link] [Span Down] is the same as [Span Down].
Command: FSPN (?) d {, i}
[Marker Ref]
[Marker Ref] toggles the Marker Rel Mode (in the [Marker] menu) between Off and
Relative for the active display. If Marker Rel is Off, [Marker Ref] sets Marker Rel to
Relative and sets the X-Rel and Y-Rel Marker Offsets to the current Marker Position.
Subsequent marker readings are relative to this offset. The relative Marker Position
values are preceded by a
∆
(delta) symbol. The Marker Offset location within the display
is marked by small flag shaped symbols.
If Marker Rel is Relative, [Marker Ref] sets Marker Rel to Off and the Marker Position
is absolute.
[Marker Ref] has no affect if the Marker Rel Mode is not Off or Relative.
[Link] [Marker Ref] toggles the Marker Rel Mode of both displays.
Command: MRON (?) d
[Display Ref]
[Display Ref] toggles the Reference Display within the active display. If the Reference
Display is off, [Display Ref] loads the Reference Display with the current display data.
The Reference Display is graphed in half intensity. This allows comparison of new data
with the stored Reference Display. Set the Marker Relative to Reference Display (in the
[Marker] menu) to read the Marker Position relative to the Reference Display. Press
[Display Ref] again to turn off the Reference Display.
If display parameters are changed to values incompatible with the Reference Display, the
reference display will be turned off.
Changing the display scale does NOT rescale the Reference Display. This allows the live
measurement display to be offset from the Reference Display by changing either Ymax,
Ymid or Ymin in the [Display Setup] menu. To graph the Reference Display in the new
display scale, press [Alt] [Snap Ref].
[Link] [Display Ref] toggles the Reference Display of both displays.
Command: DREF d, i