Capture
1-29
SR785 Dynamic Signal Analyzer
Press
<Zoom>
There are far too many points in the buffer to graph
each one. The graph shows the envelope of the data
in this case.
Zoom in to show individual points. ‘Expand’ below
the graph indicates that the graph has been
graphically zoomed and does not show all of the
data along the X axis.
Press [1] [1] and [Enter].
Enter a zoom factor of 11 (2
11
). The display now
shows the signal sine wave clearly.
7. Unplug the signal from the Ch1 input.
The signal should disappear from the spectrum in
DisplayA.
Press
[Input]
Select the Input menu.
Press <Input Source>
Change the Input Source to measure from the
Capture buffer.
Use the knob to select (Playback) and press
[Enter].
Choose Playback instead of the analog inputs.
The measurement now takes its input from the data
stored in the Ch1 Capture buffer. The signal
reappears in the spectrum in DisplayA.
Only single channel measurements using Ch1 are
allowed in this case (there is no Ch2 data
available).
Press <Playback Config>
Enter the playback configuration submenu.
Press <Playback Length>
You can choose to playback only a portion of the
buffer if desired.
Press [1] [8] [0] [0] and [Enter].
Choose the entire buffer by setting the Playback
Length equal to the Capture Length.
Press <Playback Mode>
Playback can be 1-Shot (once through the buffer
and stop) or Circular (repeat when finished).
Use the knob to select (Circular) and press
[Enter].
Choose circular playback. The % indicator at the
top of the display shows the current progress
through the playback buffer.
uffer during playback.
DisplayB automatically pans to show the portion of
the Capture buffer at the current playback position.