1-26 Octave
Analysis
SR785 Dynamic Signal Analyzer
Press
<Measurement>
Change the Measurement of both displays.
Remember, in 1 channel analysis, both displays
measure the same input.
Use the knob to select (Oct ch2) and press
[Enter].
Both displays now show the filter output on Ch2.
Press
<Measurement>
Change the Measurement again.
Use the knob to select (Oct ch1) and press
[Enter].
Show the source output on Ch1 on both displays.
9. Press [Input]
Select the Input menu.
Press <Input Config>
Show the Input Configuration submenu.
Press <Ch1 A-Wt Filter>
Choose the Ch1 Input A-Weighting filter.
Use the knob to select (On) and press [Enter].
Turn the A-Wt filter On. The Ch1 Awt indicator at
the top of the screen is highlighted.
The hardware A-Wt filter conforms to the ANSI
standard and is commonly used in sound
measurements. This filter attenuates high and low
frequencies according to how people hear and
perceive sound.
Press [Link] [Auto Scale A]
The A-Weighted spectrum is a bandpass centered
around 2 kHz.
This concludes this measurement example. You
should have a feeling for Octave measurements and
how they are setup.