SSTM (?) d {, x} <ks, [s], ms>
The SSTM command sets (queries) the Swept Sine Settle Time for display d.
The parameter x is a settle time from 0.008 to 1000 seconds. The set command
requires d=2 (both displays).
This command is valid only when the Measurement Group is Swept Sine. The
set command requires a display to be Live.
SSCY (?) d {, i}
The SSCY command sets (queries) the Swept Sine Settle Cycles for display d.
The parameter i is a number of cycles from 1 to 32767 seconds. The set
command requires d=2 (both displays).
This command is valid only when the Measurement Group is Swept Sine. The
set command requires a display to be Live.
SITM (?) d {, x} <ks, [s], ms>
The SITM command sets (queries) the Swept Sine Integration Time for display
d. The parameter x is an integration time from 0.016 to 1000 seconds. The set
command requires d=2 (both displays).
This command is valid only when the Measurement Group is Swept Sine. The
set command requires a display to be Live.
SICY (?) d {, i}
The SICY command sets (queries) the Swept Sine Integration Cycles for display
d. The parameter i is a number of cycles from 1 to 32767 seconds. The set
command requires d=2 (both displays).
This command is valid only when the Measurement Group is Swept Sine. The
set command requires a display to be Live.