4-202 System Diagnostics Menu
SR785 Dynamic Signal Analyzer
Press <Return> for the <Diagnostics> menu.
Serial Number
Reset the internally stored serial number after replacing the DSP board. The serial
number which is reported on the power-on screen and in the GPIB identification string is
stored in the calibration ROM on the DSP board.
Do not use this function unless the DSP board has been replaced!
After replacing the DSP board, enter the unit serial number (shown on the rear panel)
and use <Program S/N> to store the serial number permanently.
Press <Return> for the <Diagnostics> menu.
Program S/N
Reset the internally stored serial number after replacing the DSP board. The serial
number which is reported on the power-on screen and in the GPIB identification string is
stored in the calibration ROM on the DSP board.
Do not use this function unless the DSP board has been replaced!
After replacing the DSP board, use <Serial Number> to enter the unit serial number
(shown on the rear panel) and use <Program S/N> to store the serial number
permanently.
Press <Return> for the <Diagnostics> menu.
Kill AutoOff
Clear the internal input offset calibrations and disable Auto Offset.
Do not use this function unless you are making input offset or CMR adjustments to
an analog input board. See the Service Manual for more information.
The unit must be turned off and back on after using this function to restore the input
calibrations!