Waterfall
Display
1-35
SR785 Dynamic Signal Analyzer
Use the knob to select (Pink) and press [Enter].
Choose Pink noise. Pink noise rolls off at -3dB per
octave. This maintains equal power per octave band
and yields a flat octave spectrum.
Press
<Burst>
Change the Burst percentage.
Press [9] [6] [Enter]
Using a percentage less than 100% makes the noise
source a triggered source. 100% burst outputs noise
continuously with 100% duty cycle. Bursts less
than 100% will output noise with less than 100%
duty cycle and may be triggered.
In this case, the noise will be output for 96% of the
source period (100 ms default) or 96 ms.
If the measurement is continuously or source
triggered, then the output is noise for 96 ms out of
every 100 ms.
If the measurement is externally or manually
triggered, the output is 96 ms every trigger with the
minimum trigger period equal to 100 ms (the source
period).
6. Press [Average]
Select the Average menu. Note that this menu is
changed in Octave group.
Press <Averaging Type>
Change the Averaging Type.
Use the knob to select (Linear Time) and press
[Enter].
Use Linear Time for best time resolution.
Exponential Time averaging takes about 5
integration times to fully respond to a transient.
Linear Time averaging responds in a single
integration time.
Press <Integration Time>
Change the Integration Time.
Use the knob to select (8 ms) and press [Enter].
Choose the minimum time for the best resolution.
Press <Linear Avg Trig>
Change the Linear Average Trig. This determines
how measurements behave when triggering is
enabled (not Free Run).