3-20 Status
Indicators
SR785 Dynamic Signal Analyzer
This indicator is replaced by the Capture/Playback Progress indicator when capture
(playback) is in progress.
This indicator is replaced by the Sweep Frequency indicator when the Measurement
Group is Swept Sine.
This indicator is replaced by the RPM indicator when Show Tach is set to On in the
[input]<Tach Input> submenu.
Sweep Frequency
The Sweep Frequency indicator shows the current frequency during Swept Sine sweeps.
‘Ramping’ is displayed if the source level is currently ramping. This indicator replaces
the Realtime indicator when the Measurement Group is Swept Sine.
RPM
The RPM indicator displays the rotational speed of the device under test based on
information from the tachometer input. The rotational speed is calculated from the
freqeuency of pulses arriving at the tachometer input and the setting of the Pulses per
Revolution softkey. This indicator replaces the Real time indicator when the
Measurement Group is Order or when Show Tach is set to On.
Capture Progress
During capture fill, the Capture Progress indicator shows how much of the capture buffer
has been filled. During capture playback, this indicator shows the playback progress
through the playback portion of the buffer. This indicator replaces the Real time
indicator when capture or playback is in progress.
No Cap./Cap. Data
The Capture indicator shows the status of the Capture buffer.
‘No Cap.’ indicates that the Capture buffer is empty. Input playback from Capture is not
allowed. Arbitrary source playback from Capture is also not allowed.
‘Capturing’ indicates that capture is in progress. The capture progress is shown in the
Real Time indicator position.
‘Cap. Data’ indicates that the Capture buffer contains data for playback.
Link
When an unlinked display parameter is being entered or modified, the Link indicator
(next to the top of the menu) reads either ‘DispA’ or ‘DispB’, whichever is the active
display. Parameter entry or selection modifies only the active display. To change which
display will be modified, press [Active Display].
When a linked parameter is highlighted, the Link indicator reads ‘Link’. Parameter entry
or selection modifies both displays at once.
To link or unlink a parameter, highlight the parameter softkey and press the [Link] key.
If the Link indicator is shown in gray, then the parameter linking may not be changed.
Содержание SR785
Страница 4: ...ii ...
Страница 10: ...viii ...
Страница 80: ...1 64 Exceedance Statistics ...
Страница 158: ...2 78 Curve Fitting and Synthesis SR785 Dynamic Signal Analyzer ...
Страница 536: ...5 136 Example Program SR785 Dynamic Signal Analyzer ...