Display Setup Commands 5-37
SR785 Dynamic Signal Analyzer
Display Setup Commands
MGRP (?) d {, i}
The MGRP command sets (queries) the Measurement Group of display d. The
parameter i selects FFT (0), Correlation (1), Octave (2), Swept Sine (3), Order
(4), ord Time/Histogram (5)..
Changing the Measurement Group changes the Frequency, Display Setup and
Average menus. In addition, the Source and Capture menus may also change.
The Measurement Group specific settings in these menus change to those last
used with the new group. Parameters within these menus may only be changed
with commands which are valid within the new group.
The set command requires d=2 (both displays).
MEAS (?) d {, i}
The MEAS command sets (queries) the Measurement of display d. The
parameter i selects the measurement from the list below. Only those
measurements available in the current Measurement Group are allowed.
Each measurement has an associated view. Changing the Measurement changes
the View to the view last used with the new Measurement.
The set command requires display d to be Live.
FFT
Group
i Measurement
0
FFT
1
1
FFT
2
2
Power Spectrum 1
3
Power Spectrum 2
4
Time
1
5
Time
2
6
Windowed Time 1
7
Windowed Time 2
8
Orbit
9
Coherence
10
Cross
Spectrum
11
Frequency
Response
12
Capture Buffer 1
13
Capture Buffer 2
14
FFT User Function 1
15
FFT User Function 2
16
FFT User Function 3
17
FFT User Function 4
18
FFT User Function 5