Limit
Testing
1-59
Press
<Return>
Return to the Limit Testing menu.
Press <Limit Beep>
Select Limit Beep.
Use the knob to select (On) and press [Enter].
On enables the audible alarm. This alarm alerts you
to limit test failures.
7. Press [Display Setup]
Select the Display Setup menu.
Press
<Measurement>
Change the Measurement.
The limit segments are defined for the current
measurement, view and units. Changing any of
these parameters turns limit testing off.
The Limit Testing and Beep are turned off since the
limit segments we defined have no meaning for this
measurement.
Use the knob to select (Time1) and press
[Enter].
Change the Measurement to Time Record Ch1.
If we went back to the Limit Testing menu and
tried to edit limits now, the previous limit segments
would be lost. Each display only has a single set of
limits and they are defined for a specific
measurement, view and units.
Press
<Measurement>
Change the Measurement back.
Use the knob to select (FFT ch1) and press
[Enter].
Select FFT ch1 again.
8. Press [Analysis]
Select the Analysis menu.
Press <Limit Test>
Go to the Limit Testing menu.
Press <Limit Segments>
Select Limit Segments.
Use the knob to select (Show) and press [Enter].
Show the segments. The two segments defined
earlier are still available.
This concludes this example. Limit testing is a
powerful tool for repetitive tests. In an automated
test environment, limit segments are usually
downloaded from a host computer. The SR785
performs the limit testing in real-time and the
results are queried by the host computer.
Содержание SR785
Страница 4: ...ii ...
Страница 10: ...viii ...
Страница 80: ...1 64 Exceedance Statistics ...
Страница 158: ...2 78 Curve Fitting and Synthesis SR785 Dynamic Signal Analyzer ...
Страница 536: ...5 136 Example Program SR785 Dynamic Signal Analyzer ...