4-104 Trigger Menu
SR785 Dynamic Signal Analyzer
source output. Turn the source off (or set it to Sine) when making triggered
measurements without the source.
Command: STMD (?) {i}
RPM/Time Arm Setup
Select the RPM/Time Arm menu to view and adjust parameters relating to RPM and
Time arming.
Start RPM
Specify the threshold RPM for RPM arming. Start RPM is only active if Start RPM
(On/Off) is On.
For the trigger to initially arm, the RPM must exceed the threshold RPM (if Delta RPM
Sense is Increasing), or must be below the threshold (if Delta RPM Sense is
Decreasing.). If Delta RPM sense is set to Abs. Change, the Start RPM is ignored.
Command: TRSR (?) {f}
Start RPM (On/Off)
Activates the Start RPM threshold. If Start RPM is set On, the trigger will not be armed
until the RPM passes the threshold. If Start RPM is Off, the threshold setting is ignored.
If Start RPM is On, the RPM threshold must be met for both the RPM arming and Time
Arming trigger modes.
Command: TRSM (?) {i}
Delta RPM
Set the RPM increments for which the trigger will be armed. Once the Start RPM
condition, if any, has been met. The trigger will arm each time the rpm changes by the
specified amount.