Frequency
Commands
5-29
SR785 Dynamic Signal Analyzer
This command is valid only when the Measurement Group is FFT. The set
command requires display d to be Live.
Since correlation is always a baseband measurement, this command is not valid
when the measurement group is correlation.
FEND (?) d {, f} <kHz, [Hz], mHz,
µµµµ
Hz>
The FEND command sets (queries) the FFT End frequency of display d. The
parameter f is a frequency (real number of the specified units). Values of f which
would cause the span to exceed the 0 to 102.4 (100.0) kHz range cause an error.
This command is valid only when the Measurement Group is FFT. The set
command requires display d to be Live.
Since correlation is always a baseband measurement, this command is not valid
when the measurement group is correlation.
UNST d
The UNST command unsettles the measurement of display d. The measurement
is not actually perturbed by the UNST command. The settling status is set to
unsettled and the full settling time of the measurement is required before the
status returns to settled.
The measurement is unsettled by changing any one of several measurement
parameters. For example, changing the input range or FFT span will unsettle the
measurement. If the signal comes from an external source and is changed in such
a way as to require the measurement to settle, it is convenient to use the UNST
command and wait for settling to finish.
This command is valid only when the Measurement Group is FFT or Octave.