4-40 Display Setup Menu
SR785 Dynamic Signal Analyzer
For two channel measurements, the phase is channel 2 relative to channel 1. Triggering
is not generally required for meaningful two channel phase measurements.
Phase suppress sets the phase of small data values to zero. This avoids the messy phase
display associated with the noise floor. (Remember, even a small signal has phase.) Set
the phase suppress threshold in the [Display Options] menu.
Nyquist Plot
The Nyquist plot graphs complex measurement data as imaginary part (along the Y axis)
versus real part (along the X axis). The graph is scaled linearly. Adjacent frequency
points are connected by a line. This view is generally meaningful only for measurements
which have data at every frequency point (such as chirp source or swept sine).
The marker readout shows the real and imaginary parts of each point as well as its
frequency. The marker moves sequentially through the frequency points and can be
linked to the frequency of the other display.
Real measurement data, such as baseband time record, have zero imaginary part. This
view is entirely along the X axis.
Nichols Plot
The Nichols plot graphs complex measurement data as log magnitude (along the Y axis)
versus unwrapped phase (along the X axis). The Y axis is logarithmic and the X axis is
linear. Adjacent frequency points are connected by a line. This view is generally
meaningful only for measurements which have data at every frequency point (such as
chirp source or swept sine).
The marker readout shows the log magnitude and phase of each point as well as its
frequency. The marker moves sequentially through the frequency points and can be
linked to the frequency of the other display.
Real measurement data, such as baseband time record, have zero phase. This view is
entirely along the Y axis.
Units
Display the Units submenu. Press <Return> for the main [Display Setup] menu.
Choices in the units submenu are limited to those which are appropriate for the current
Measurement and View. For instance, dB units may not be selected when the view is
Real Part. Settings in the units submenu are associated with the Measurement and View.
Changing Measurements or Views may change the settings within the Units submenu