Waterfall
Commands
5-81
SR785 Dynamic Signal Analyzer
This command is valid only when the Measurement Group is Octave.
WVCT (?) d {, i}
The WVCT command sets (queries) the Waterfall View Count for display d. The
parameter i is a number of records.
This command is not valid when the Measurement Group is Swept Sine.
WHIT (?) d {, i}
The WHIT command sets (queries) the Waterfall Trace Height for display d. The
parameter i is a percentage [20..80] of the total display height.
This command is not valid when the Measurement Group is Swept Sine.
WANG (?) d {, i}
The WANG command sets (queries) the requested Waterfall Angle for display d.
The parameter i is a signed integer number of degrees [-75..+75]. The display
will scroll as close to this angle as possible.
This command is not valid when the Measurement Group is Swept Sine.
WFST (?) d {, i}
The WFST command sets (queries) the Waterfall Fast Angles for display d. The
parameter i selects Off (0) or (On).
This command is not valid when the Measurement Group is Swept Sine.
WTHR (?) d {, i}
The WTHR command sets (queries) the Waterfall Threshold for display d. The
parameter i is a percentage of the full scale trace height.
This command is not valid when the Measurement Group is Swept Sine.
WHID (?) d {, i}
The WHID command sets (queries) the Waterfall Hidden Lines for display d.
The parameter i selects Invisible (0) or Visible (1).
This command is not valid when the Measurement Group is Swept Sine.
WREV (?) d {, i}
The WREV command sets (queries) the Waterfall Paused Drawing for display d.
The parameter i selects Normal (Newest at Top) (0) or Reversed (Oldest at Top)
(1).
This command is not valid when the Measurement Group is Swept Sine.
WTRC d, i, j
The WTRC command saves waterfall record j [0 is most recent, 1 is next, etc.]
from display d to Trace i [1..5].