4-20 Order Frequency Menu
SR785 Dynamic Signal Analyzer
Delta Order
Sets the resolution for order spectra. The value may range between Max Order divided
by 400, and 1.0. Choosing a smaller value for Delta Order (higher resolution) increases
the time it takes to acquire an order spectrum.
Command: ODLT (?) d {, f}
Tracking
Turns order tracking on and off. When tracking is On, the Track1 and Track2
measurments become available in the order measurement group. The Track
measurements are plots of the amplitude of one particular order as a function of RPM.
Because track measurements use the memory allocated to the waterfall buffer, you
cannot use waterfall storage while Tracking is on.
Command: OTRK (?) d {, i}
Track Setup
Displays the track setup submenu. This submenu displays parameters which define the
Track1 and Track2 measurements.
Track Points
Sets the number of points for track measurements. When the specified number of points
have been acquired the measurement will either stop, or wrap around depending on the
setting of the track storage softkey.
A track measurement is a two-dimensional plot of the amplitude of a particular order vs.
rpm with time as the parameter. Individual points in the measurement can be taken
continuously, or at specified time or rpm intervals using Time arming and RPM Arming.
Command: ONPT (?) d {, i}