5-32 Frequency
Commands
SR785 Dynamic Signal Analyzer
Frequency Commands (Swept Sine)
SSTR (?) d {, f} <kHz, [Hz], mHz,
µµµµ
Hz>
The SSTR command sets (queries) the swept sine Start frequency of display d.
The parameter f is a frequency (real number of specified units). The range of f is
0.001 to 102.4E3. The set command requires d=2 (both displays).
This command is valid only when the Measurement Group is Swept Sine. The
set command requires a display to be Live.
SSTP (?) d {, f} <kHz, [Hz], mHz,
µµµµ
Hz>
The SSTP command sets (queries) the swept sine Stop frequency of display d.
The parameter f is a frequency (real number of specified units). The range of f is
0.001 to 102.4E3. The set command requires d=2 (both displays).
This command is valid only when the Measurement Group is Swept Sine. The
set command requires a display to be Live.
SSFR ?
The SSFR command queries which sweep point was the most recently measured.
While a sweep is in progress, SSFR? returns values from 0 to the Number Of
Points - 1.
After a sweep is started, SSFR? will not return 0 until the first point is measured.
This may take some time if Source Ramping is On or if the Settling Time is very
long.
This command is valid only when the Measurement Group is Swept Sine.
SRPT (?) d {, i}
The SRPT command sets (queries) the swept sine Repeat Mode of display d. The
parameter i selects Single Shot (0) or Continuous (1). The set command requires
d=2 (both displays).
This command is valid only when the Measurement Group is Swept Sine. The
set command requires a display to be Live.
SSTY (?) d {, i}
The SSTY command sets (queries) the swept sine Sweep Type of display d. The
parameter i selects Linear (0) or Logarithmic (1). The set command requires d=2
(both displays).
This command is valid only when the Measurement Group is Swept Sine. The
set command requires a display to be Live.
SARS (?) d {, i}
The SARS command sets (queries) the swept sine Auto Resolution Mode of
display d. The parameter i selects Off (0) or On (1). The set command requires
d=2 (both displays).
Содержание SR785
Страница 4: ...ii ...
Страница 10: ...viii ...
Страница 80: ...1 64 Exceedance Statistics ...
Страница 158: ...2 78 Curve Fitting and Synthesis SR785 Dynamic Signal Analyzer ...
Страница 536: ...5 136 Example Program SR785 Dynamic Signal Analyzer ...