Saving and Recalling 1-47
SR785 Dynamic Signal Analyzer
Saving and Recalling
This example illustrates saving and recalling displays to reference displays, traces and disk files.
1. Press [System]
Display the System menu.
Press
<Preset>
Preset returns the unit to its default settings.
Press [Enter] to confirm Preset.
Preset requires confirmation to prevent accidental
reset. Wait until the self tests are completed.
2. Connect the Source Output to the Channel 1 A
Input.
Setup to analyze the source output. The default
source is a 10.24 kHz sine.
Press [Source], <On>
Turn the source on.
Press [Span Down] twice to change the span to
25.6 kHz.
Narrow the span to display the signal better.
Press [Auto Scale A]
Scale the display to show the entire measurement
range.
3. Press [Display Ref]
Copy the current measurement data into the
reference graph for the active display (A).
The reference graph is stored data which is
associated with each display. The reference graph
is shown in half intensity.
We need to change the measurement data in order
to see the reference graph since it is ‘underneath’
the current data.
Press [Input], <Input Conifg>
Select the Input Configuration submenu.
Press <Ch1 Input Range>
Change the input range to raise the noise floor.
Press [3] [0], select (dbVpk) with the knob, and
[Enter].
Select 30 dBV for the Ch1 Input Range.
Now you can see the reference graph below the
current measurement’s noise floor. The reference
graph allows visual comparison of live data with
stored data. The marker can also be set to read the
current data relative to the reference graph.
The reference graph can be loaded by copying the
current live data or by copying a stored trace.