The Measurement Group determines which Average menu is displayed. This menu sets
the averaging parameters for the measurement. A set of parameters is stored for each
group. No averaging parameters are available in the Time/Histogram measurement
group.
FFT, Correlation, Order Average Menu
When the Measurement Group is FFT, Correlation, or Order, this menu sets the
averaging parameters for the active display. If the Analyzer Configuration is Dual
Channel, the averaging parameters are the same for both displays. Averaging successive
measurements together improves accuracy and repeatability.
Compute Averages
Turns the computation of averages on and off. If Compute Averages is off, no averaged
quantities will be computed or displayed. Although this results in a slight improvement
in the speed of some measurements, for normal operation Compute Averages should be
left On.
Some FFT measurements, such as coherence, have averaging as part of their definition.
If one of these measurements is selected, Compute Averages will be forced On.
Command: FAVG (?) d {, i}