4-58 Harmonic Marker Menu
SR785 Dynamic Signal Analyzer
Harmonic Marker Menu
This menu is displayed when the active display Marker Mode is Harmonic.
Harmonics
Set the Number Of Harmonics for the active display [1.. 100]. The second harmonic
(2xfundamental) is identified as #1, etc.
The Harmonic Marker identifies the fundamental frequency with the Fundamental
Marker (solid vertical line) and the number of harmonics with small triangular Harmonic
Markers. The Marker Position Display can read the Fundamental or a single Harmonic.
Only those harmonics within the measurement span which are identified by the
Harmonic Markers contribute to the THD calculations. The harmonic identification
works best if the fundamental frequency is a bin frequency. If the fundamental is off-bin,
then it is best to use the Flattop window to broaden all of the peaks.
Each display has its own Number Of Harmonics. This entry field can be linked to both
displays using the [Link] key.
Command: HRMN (?) d {, i}
Display
Select the Harmonic and Sideband Display for the active display. The Marker Position
can display the Fundamental or any identified Harmonic or Sideband.
For Harmonic Marker, the minimum is 0 (Fundamental) and the maximum is the Number
Of Harmonics. The second harmonic (2xfundamental) is identified as #1, etc.