1-16 Linking (Advanced Operation)
SR785 Dynamic Signal Analyzer
Select (Independent Chan.)
In this example we'll want to set separate frequency
spans for the two displays. To do that the analyzer
must be in the Independent Channels mode. In the
default Dual Channel mode the span is the same for
both channels, but dual channel measurements are
allowed.
7. Press [Freq]
Select the Frequency menu. The menu shows the
frequency parameters for the measurement in
DisplayA (active display).
Press
<Span>
Highlight the Span. Note that the Link indicator at
the top of the screen turns on. This indicates that
the highlighted parameter (Span) is linked to both
displays. Changing a linked parameter affects both
displays at once.
Use the knob to adjust the Span to 12.8 kHz and
press [Enter].
Narrow the Span of both displays to show the filter
notch at 1 kHz (noisy of course).
Press <Span> again.
Highlight the Span again.
Press
[Link]
Pressing [Link] toggles parameter linking off. The
Link indicator now shows ‘DispA’ indicating that
this menu box adjusts the span for DisplayA only.
Use the knob to adjust the Span to 3.2 kHz and
press [Enter].
Change the Span of DisplayA to 3.2 kHz. The Span
of DisplayB remains at 12.8 kHz. The SR785
allows the two displays to have differing Spans and
Start frequencies in the Independent Channels
mode.
Many parameters affect the displays separately.
Linking is a convenient way to adjust the two
displays together and keep their settings the same.
The default settings link many measurement
parameters, such as frequency and averaging,
together as found in many other instruments.
8. Press [Average]
Select the Average menu.
Press <Display Avg>
Select the averaging mode to display. The SR785
always computes all the averaging modes when
<Compute Avgs> is set to Yes. The <Display Avg>
softkey selects which averaging mode is currently
displayed.
Select (RMS) with the knob and press [Enter].
Select RMS Averaging for both displays.