4-194 System Remote Menu
SR785 Dynamic Signal Analyzer
System Remote Menu
Output To
Select the Output Interface [RS232, GPIB]. All interface query responses are directed to
the selected Output Interface regardless of which interface received the query. Use the
OUTX command to select the correct interface at the beginning of every program.
Command: OUTX (?) {i}
GPIB Address
Enter or adjust the instrument GPIB Address [0..30].
Overide REM?
Select the Overide Remote mode [Yes, No]. In general, every GPIB interface command
will put the unit into the REMOTE state with the front panel locked out. To defeat this
feature, set this entry to Yes. In this mode, the front panel is not locked out when the unit
is in the REMOTE state.
To return from REMOTE to LOCAL (front panel enabled), press [Help/Local].
Command: OVRM (?) {i}
Baud Rate
Select the RS232 (Serial) interface Baud Rate [300, 1200, 2400, 4800, 9600, 19.2k].
Most PC’s use 9600 as a default.