3-18 Status
Indicators
SR785 Dynamic Signal Analyzer
Status Indicators
In addition to the data displays and menu boxes, there are a number of status indicators
which are displayed at the top of the screen.
Figure Chapter 3 -9 Status Indicator Panel
Input Ranges
The Input Ranges for both inputs are displayed. If the range is set manually, the range
indicator is on a normal background. If Auto Range is On, then the range is displayed on
a highlighted background. The input range is displayed in whatever units it was last
entered in with the menu.
Overloads
The input overload condition of the inputs are displayed to the right of their Input
Ranges. The overload indicator is a bar. When the bar is 1 division long (displayed
faintly), then the input signal is below half full scale. When the bar is 2 divisions long
(displayed in black), the input signal is above half full scale (but not overloaded). When
the bar is 4 divisions long (displayed brightly), the input signal exceeds full scale and is
overloaded. This also sounds the audible overload alarm.
Once an overload occurs, the outline of the overload bar remains in the indicator. The
overload outline indicates that the input has overloaded since the measurement was
started. This is important when the measurement is linearly averaged and an overload
occurs during the average. Pressing [Start/Reset] restarts the measurement and erases this
overload outline.
Whenever an input greater than 57 V is detected at the input, the overload indicator
displays ‘HighV’ on a bright background. The analyzer immediately switches to 34 dBV
range to protect the input circuitry. Any attempt to change the input will not be allowed
until the input signal is reduced to a safe level.
Input Configurations
The Input Mode and Coupling for both inputs are displayed. If the coupling is DC or AC,
the coupling indicator is on a normal background. If ICP coupling is selected, then ICP is
displayed on a highlighted background.
AWt
The A-Weighting filter status for both inputs are displayed. If the A-Weighting filter is
ON, then ‘AWt’ is displayed on a highlighted background.
ArmWait/Trig-wait/Trig/Acquire
This indicator reads ‘Trig-wait’ while waiting for a trigger. Once triggered, the indicator
flashes ‘Trig’ and then displays ‘Acquire’ while the time record is being acquired. If the