5-60 Input
Commands
SR785 Dynamic Signal Analyzer
Input Commands
ISRC (?) {i}
The ISRC command sets (queries) the Input Source. The parameter i selects
Analog (0) or Capture (1).
Capture is not valid when the Measurement Group is Swept Sine.
LINK (?) {i}
The LINK command sets (queries) the Analyzer Configuration. The parameter i
selects Independent Channels (0) or Dual Channels (1).
I1MD (?) {i}
The I1MD command sets (queries) the Ch1 Input Mode. The parameter i selects
A (single-ended) (0) or A-B (differential) (1).
I1GD (?) {i}
The I1GD command sets (queries) the Ch1 Input Grounding. The parameter i
selects Float (0) or Ground (1).
I1CP (?) {i}
The I1CP command sets (queries) the Ch1 Input Coupling. The parameter i
selects DC (0), AC (1) or ICP (2).
I1RG (?) {x} <[dBVpk], dBVpp, dBVrms, Vpk, Vpp, Vrms, dBEUpk, dBEUpp, dBEUrms, EUpk,
EUpp,EUrms>
The I1RG command sets (queries) the Ch1 Input Range. The parameter x is the
full scale input range in the specified untis. If the exact input range specfied is
not available, the closest available input range will be used.
The query command returns a value of the form x,j where x is the value and j is
the index into the list of units.
If Ch1 AutoRange is On, the I1RG i command will turn Ch1 AutoRange Off and
the set the Ch1 Range to f.
A1RG (?) {i}
The A1RG command sets (queries) the Ch1 AutoRanging Off/On. The
parameter i selects Off (Manual) (0) or On (AutoRanging) (1). If i=1 and Ch1
AutoRange is already On, a new AutoRange is performed.
I1AR (?) {i}
The I1AR command sets (queries) the Ch1 AutoRange Mode. The parameter i
selects Up Only (0) or Tracking (1).
I1AF (?) {i}
The I1AF command sets (queries) the Ch1 Anti-Aliasing Filter Off/On. The
parameter i selects Off (0) or On (1).
Содержание SR785
Страница 4: ...ii ...
Страница 10: ...viii ...
Страница 80: ...1 64 Exceedance Statistics ...
Страница 158: ...2 78 Curve Fitting and Synthesis SR785 Dynamic Signal Analyzer ...
Страница 536: ...5 136 Example Program SR785 Dynamic Signal Analyzer ...