4-10 FFT Frequency Menu
SR785 Dynamic Signal Analyzer
to the Linewidth. If an entered Start frequency is incompatible with the Span, then it will
be set to the closest allowed value.
This softkey also anchors the Start frequency. Adjusting the FFT Span will leave the
Start frequency unchanged. Enlarging the frequency Span may change the Start
frequency if the new measurement would extend below 0 Hz or above 102.4
(100.0) kHz.
If the Measurement Group is Correlation, then the Start frequency is always 0 Hz
(baseband).
The two displays can have different Start Frequencies if the Analyzer Configuration is
set to Independent Channels. In this configuration, no two channel measurements are
allowed (frequency response, cross-spectrum, etc.), but the entry field can be linked or
unlinked using the [Link] key. If Analyzer Configuration is set to Dual Channel they
field is automatically linked to both displays.
Command: FSTR (?) d {, f}
Center Frequency
Set the Center frequency of the active display FFT span. The Center frequency is the
center of the measurement span. The knob adjusts the Center frequency in steps equal to
the Linewidth. If an entered Center frequency is incompatible with the Span, then it will
be set to the closest allowed value.
This softkey also anchors the Center frequency. Adjusting the FFT Span will leave the
Center frequency unchanged. Enlarging the frequency Span may change the Center
frequency if the new measurement would extend below 0 Hz or above 102.4
(100.0) kHz.
If the Measurement Group is Correlation, then the Center frequency is always equal to
Span/2 (baseband).
The two displays can have different Center Frequencies if the Analyzer Configuration is
set to Independent Channels. In this configuration, no two channel measurements are
allowed (frequency response, cross-spectrum, etc.), but the entry field can be linked or
unlinked using the [Link] key. If Analyzer Configuration is set to Dual Channel they
field is automatically linked to both displays.
Command: FCTR (?) d {, f}
End Frequency
Set the End frequency of the active display FFT span. The End frequency is the highest
frequency in the measurement span. The knob adjusts the End frequency in steps equal to
the Linewidth. If an entered End frequency is incompatible with the Span, then it will be
set to the closest allowed value.