4-46 Display Options Menu
SR785 Dynamic Signal Analyzer
Display Options Menu
The Display Options menu configures the active display format.
Display
Select the Display Update of the active display [Live, Off-Line].
A Live display updates continuously to show live measurement results (as long as the
measurement is not paused).
An Off-Line display does not update. The display shows a ‘frozen’ measurement result
or trace. A display becomes Off-Line whenever a trace is recalled to the display from
disk or memory. Since an Off-Line display is showing old data, its Measurement,
Window, Frequency Span and Averaging may not be modified. Only display related
parameters such as View, Scaling and Marker Functions may be changed Off-Line.
The two displays can have different Display Update Modes. This entry field can be
linked to both displays using the [Link] key.
A single display may be taken Off-Line while the other display is still Live. This allows
comparison of live results with a previous result. This is unlike the [Pause/Cont] key
which pauses ALL measurements.
Command: DISP (?) d {, i}
Format
Select the screen Display Format [Single, Dual, or Overlay].