Sine Source Menu 4-71
SR785 Dynamic Signal Analyzer
Sine Source Menu
The Sine Source menu is used to set the frequency, offset, and amplitude of the sine
source.
Frequency 1
Set the Frequency of Tone 1. The sine output is the sum of two tones (sine waves). When
the knob is used to adjust the frequency, the resolution is equal to the Linewidth of the
active display FFT Span. The knob always sets the frequency to an exact multiple of the
Linewidth. For octave analysis, the Linewidth of the most recent FFT Spans are used
(with an FFT Base of 100.0 kHz).
The keypad allows arbitrary frequencies to be entered. Remember, the output is periodic
over the FFT time record only if the frequency is an exact multiple of the Linewidth.
Source Trigger will not result in a stable phase for non-periodic frequencies.
The sine source is not triggered. The output is always continuous.
Command: S1FR (?) {f}
Amplitude 1
Set the peak Amplitude of Tone 1 [0..5V]. The amplitude resolution is 0.1 mV. The sine
output is the sum of two tones (sine waves) and the offset (constant). To generate a
single tone, set the amplitude of one of the tones to zero. Note that the sum of the
amplitudes of Tone 1, Tone 2 and the absolute value of the offset cannot exceed 5 V.
Command: S1AM (?) {x}