4-88 Input
Menu
SR785 Dynamic Signal Anaylzer
When the measurement group is Swept Sine or Order, the Analyzer Configuration is
always set to Dual Channel.
Command: LINK (?) {i}
Auto Offset
Enable Auto Offset input calibration [Off, On].
When Auto Offset is On, the analyzer will periodically perform an input offset
calibration. This involves internally grounding both inputs and correcting their DC
offsets. This only takes a second. However, the measurements will need to settle again
after an offset calibration. To avoid disrupting a lengthy measurement, turn Auto Offset
Off.
A warning message is displayed 15 seconds before an Auto Offset calibration is
performed. Pressing [<-] (backspace) will abort the scheduled calibration.
Command: IAOM (?) {i}