Waterfall
Display
1-39
SR785 Dynamic Signal Analyzer
Move the marker all the way to the right hand
edge of the display. The marker position should
read ‘L:49’.
The last bin in the octave display is the total sound
level L
eq
(as selected by <Power Bin>).
12. Press [Waterfall]
Select the Waterfall menu.
Press
<More>
Show More of the Waterfall menu.
Press <Slice to Trace>, select (Trace 1) with the
knob and press [Enter].
A waterfall slice is the time history of a single X
axis location (data at the marker X position from all
stored records). The slice data is stored in a trace.
Press [Active Display]
Make the bottom display active (DisplayB).
Press [Alt] [Help/Local], select (Trace 1) with
the knob and press [Enter].
Trace to Display is an alternate function. The
alternate key functions are labeled below each key
(in this case, the [Help/Local] key).
Trace to Display recalls trace data to the active
display.
Press [Auto Scale B]
Auto Scale DisplayB to show the time history of
Leq.
The points in a slice are numbered and displayed
from 0 (oldest) to 49 (newest). Note that this differs
from the waterfall display in which the newest
record is numbered 0. This is because the slice is a
time record with time advancing to the right and it
is more natural to number it this way.
Use the knob to move the marker around in the
waterfall slice (DisplayB).
Note that the first 12 records (96 ms) show a large
value for L
eq
during the noise burst. If this was a
real reverberation measurement, the signal would
not decay in a single 8 ms measurement but would
last for a reverberation time.
13. Press [Active Display]
Make the top display active (DisplayA).
Press
[Waterfall]
Select the Waterfall menu.
Press <View Count>
Change the View Count. This is the number of
records which are displayed.
Press [5] [0] [Enter].
Enter 50 to show the entire waterfall buffer.
Press <Marker Z to>
Move the marker to a specific record number.
Содержание SR785
Страница 4: ...ii ...
Страница 10: ...viii ...
Страница 80: ...1 64 Exceedance Statistics ...
Страница 158: ...2 78 Curve Fitting and Synthesis SR785 Dynamic Signal Analyzer ...
Страница 536: ...5 136 Example Program SR785 Dynamic Signal Analyzer ...