5-50 Marker
Commands
SR785 Dynamic Signal Analyzer
Marker Commands (Sideband)
SSEP (?) d {, f} <kHz, [Hz], mHz,
µµµµ
Hz>
The SSEP command sets (queries) the Sideband Separation for display d. The
parameter f is the sideband separation in Hz.
This command is only valid if the Marker Mode for display d is Sideband.
SDBN (?) d {, i}
The SDBN command sets (queries) the Number of Sidebands for display d. The
parameter i is a number of sidebands from 1 to 50.
This command is only valid if the Marker Mode for display d is Sideband.
HDSP (?) d {, i}
The HDSP command sets (queries) the Harmonic and Sideband Marker Display
for display d.
For Sideband Marker, i=0 selects the Fundamental, i=-1 selects the first lower
sideband, i=1 selects the first upper sideband. The maximum value of i plus or
minus the Number of Sidebands. i<0 is valid only for Sideband marker.
This command is only valid if the Marker Mode for display d is Harmonic or
Sideband.
HRDO (?) d {, i}
The HRDO command sets (queries) the Harmonic and Sideband Marker Readout
for display d. The parameter i selects Absolute (0) or Relative to Fundamental
(1).
This command is only valid if the Marker Mode for display d is Harmonic or
Sideband.
SPWR ? d, i
The SPWR ? command queries the Sideband Power for display d. The parameter
i selects dB Relative to Fundamental (0) or Vrms or dBVrms (1) and is required.
This command is only valid if the Marker Mode for display d is Sideband.