5-126 Status Word Definitions
SR785 Dynamic Signal Analyzer
Standard Event Status Word
Bit Name Set
when
0 unused
1 unused
2
QRY
Too many responses are pending
3
DDE
Too many commands received at once
4
EXE
Command cannot execute successfully
5
CME
Command syntax error
6
URQ
A key is pressed or the knob rotated
7
PON
Power is turned on
8-15
unused
reserved by IEEE standard
The Standard Event status word is defined by IEEE-488.2 (1987) and is used
primarily to report errors in commands received over the interface. These status
bits remain set until read using *ESR?. They are also cleared by the *CLS
command. Use *ESE to set bits in the Standard Event enable register.
Instrument Status Word
Bit Name Set
when
0
TRIG
A measurement has been triggered
1
DISK
A disk operation has been completed
2
OUTP
A hardcopy output operation has completed
3
TACH
A tach pulse has been received
4
CAPT
The capture buffer has filled
5
PAUS
Measurement has been paused
6
STRT
Measurement has been started
7
PLBK
Single Shot Capture Playback has finished
8
PREV
Measurement stopped to wait for average preview
9-15 unused
The Instrument status word reports on the overall measurement status of the
instrument. These status bits remain set until read by INST?. They are also
cleared by the *CLS command. Use INSE to set bits in the Instrument status
enable register.