Remote
Programming
5-27
SR785 Dynamic Signal Analyzer
Things to Remember!
Output Interface (RS232 or GPIB)
All responses are directed to the interface selected by <Output To> in the [System]
<Remote> menu, regardless of which interface received the query. Use the OUTX
command to select the correct interface at the beginning of every program.
Active Display and Linking
Set commands which require a 'd' parameter specify DisplayA (0), DisplayB (1) or Both
Displays (2) without regard to the active display or any parameter linking. Active display
and linking are screen features which do not apply to interface commands.
Analyzer Configuration and Linking
When the Analyzer Configuration is set to Dual Channel, many commands that have a 'd'
(display) parameter require d be set to '2' (both channels). This reflects the fact that the
two displays cannot have different frequency parameters in the dual channel
configuration.
Command Context Errors
Commands which perform softkey functions are not allowed unless the corresponding
softkey function is allowed. If the softkey is not available, either shown in gray or not
displayed at all, then the corresponding set command is not allowed.
Be careful to send commands in the correct order to avoid context errors. A good
practice is to send the commands in the same order as programming the instrument using
the softkeys.