Octave Average Menu 4-115
SR785 Dynamic Signal Analyzer
Octave Average Menu
When the Measurement Group is Octave Analysis, this menu sets the averaging
parameters for the active display.
Averaging Type
Select the Averaging Type for the active display [Linear Time, Exponential Time, Peak
Hold, Equal Confidence].
Choosing Peak Hold also sets the Power Bin to Peak.
Octave measurements are always rms averaged to measure the power in each band.
The [Start/Reset] key resets the current averages and starts the measurement over. The
[Pause/Cont] key pauses the measurement. Pressing [Pause/Cont] again resets the
averages and starts the measurement over.
The Averaging Type of each display can be selected separately. This entry field can be
linked to both displays by using the [Link] key.
Command: OTYP (?) d {, i}
Linear Time
The band filter outputs are equally weighted and averaged for an Integration Time.
While Linear averaging is in progress, the integration time completed is shown in the
Horizontal Scale Bar below the graph. When the Integration Time has been completed,
the measurement stops and ‘Done’ is displayed below the graph. If the ‘ Done’ alarm is
enabled, an audible alert is also sounded.
Waterfall Storage
If Waterfall Storage is On, the waterfall buffer only stores the completed linear averages,
not each individual measurement.
When the measurement is free running (Auto Arm, Continuous Trigger), each time the
linear average is done, the result is stored in the waterfall buffer and the average is reset
and started over (instead of stopping). Each completed average counts as a single
waterfall record.