Swept Sine Frequency Menu 4-15
SR785 Dynamic Signal Analyzer
Swept Sine Frequency Menu
When the Measurement Group is Swept Sine, this menu sets the frequency sweep
parameters. These parameters govern the measurements on both displays. See ‘Swept
Sine’ in Chapter 2 for a discussion about swept sine measurement fundamentals.
Start
Set the Start frequency [1 mHz .. 102.4 kHz]. Note that measurements at frequencies less
than 1 Hz take a significant amount of time.
The Start frequency is usually less than the Stop frequency. If the Start frequency is
greater than the Stop frequency, then the sweep proceeds downward from the Start
frequency.
If the Start frequency is changed during a sweep, the sweep will be reset.
Command: SSTR d (,?) {f}
Stop
Set the Stop frequency [1 mHz .. 102.4 kHz]. Note that measurements at frequencies less
than 1 Hz take a significant amount of time.
The Stop frequency is usually greater than the Start frequency. If the Stop frequency is
less than the Start frequency, then the sweep is downward from the Start frequency.
If the Stop frequency is changed during a sweep, the sweep will be reset.
Command: SSTP (?) d {, f}