Front Panel Commands 5-103
SR785 Dynamic Signal Analyzer
RCTR d, i
The RCTR command recalls Trace i to display d. If Trace i does not have data,
then an error occurs.
SVRF d, i
The SVRF command saves the Reference Display of display d to Trace i.
RCRF d, i
The RCRF command recalls Trace i to the Reference Display of display d. If
Trace i does not have data or is not compatible with the active display
measurement, an error occurs. For example, if Trace i is a time record, it cannot
be recalled into the Reference Display of an FFT measurement.
PRNT
The PRNT command prints the screen using the selected Printer Type and
Destination. All other operations are disabled until printing is completed.
PLOT
The PLOT command plots the screen using the selected Plotter Type and
Destination. All other operations are disabled until plotting is completed.
DUMP
The DUMP command dumps the data in the active display in ASCII format to
the selected Destination (Interface or Disk). The data is written f,y,x[cr] with the
data for each point on a single line. f is the frequency or time value, y is the y-
axis value and x is the x-axis value (if necessary)
ASCL d
The ASCL command AutoScales display d. This is the same as pressing the
[Auto Scale A] or [Auto Scale B] keys. The displays are AutoScaled separately
(d=2 is not allowed).
A1RG (?) {i}
The A1RG command sets (queries) the Ch1 AutoRanging Off/On. The
parameter i selects Off (Manual) (0) or On (AutoRanging) (1). If i=1 and Ch1
AutoRange is already On, a new AutoRange is performed.
A2RG (?) {i}
The A2RG command sets (queries) the Ch2 AutoRanging Off/On. The
parameter i selects Off (Manual) (0) or On (AutoRanging) (1). If i=1 and Ch2
AutoRange is already On, a new AutoRange is performed.
MRON d
The MRON command sets the Normal Marker offsets (X and Y) to the current
marker position and sets the Marker Relative to Offset (if Marker Rel was Off).
This is similar to the [Marker Ref] key.
This command is only valid if the Marker Mode for display d is Normal.
Содержание SR785
Страница 4: ...ii ...
Страница 10: ...viii ...
Страница 80: ...1 64 Exceedance Statistics ...
Страница 158: ...2 78 Curve Fitting and Synthesis SR785 Dynamic Signal Analyzer ...
Страница 536: ...5 136 Example Program SR785 Dynamic Signal Analyzer ...