Source
Commands
5-55
SR785 Dynamic Signal Analyzer
Chirp Source Commands
CAMP (?) {x} <mV, [V}, dBVpk>
The CAMP command sets (queries) the Chirp Amplitude. The parameter x is the
amplitude in the specified units. The query returns two numbers of the form y,i
where y is a real number and i is an index indicating the units.
This command is valid only when the Source Type is Chirp.
CBUR (?) {x}
The CBUR command sets (queries) the Chirp Burst Percentage. The parameter x
is a percentage from 1 to 100.
This command is valid only when the Source Type is Chirp and the
Measurement Group is FFT or Correlation.
CSRC (?) {i}
The CSRC command sets (queries) the Source Display. The parameter i selects
DisplayA (0) or DisplayB (1)
This command is valid only when the Source Type is Chirp or Noise and the
Measurement Group is FFT or Correlation.