Commands
5-59
SR785 Dynamic Signal Analyzer
SSLL (?) {x}
The SSLL command sets (queries) the Swept Sine Reference Lower Limit. The
parameter x is a ratio in dB from -0.1 to -30.0 dB.
This command is valid only when the Measurement Group is Swept Sine and
Auto Level Reference is Ch1 or Ch2.
SMAX (?) {x} <mV, [V[, dBVpk>
The SMAX command sets (queries) the Swept Sine Maximum Level. The
parameter x is the maximum level in the specified units. The query returns two
numbers of the form y,i where y is a real number and i is an index indicating the
units.
This command is valid only when the Measurement Group is Swept Sine and
Auto Level Reference is Ch1 or Ch2.
SOFF (?) {x} <mV, [V]>
The SOFF command sets (queries) the Offset of the Swept Sine Source. The
parameter x is a real number of the specified units. The query returns two
numbers of the form y,i where y is a real number and i is an index indicating the
units.
This command is valid in the swept sine measurement group.