The Arbitrary Source menu is used to configure the arbitrary source.
Amplitude
Set the Amplitude of the arbitrary source relative to 1V [0%..500%]. The maximum
output is 5V.
If the Arbitrary Waveform has been copied from a Trace, the data is normalized so that
the point with the largest absolute value is output at
±
1V when the Amplitude is 100%.
When the Arbitrary Waveform is loaded from the computer interface, the data ranges
from -1.0 to +1.0. A value of 1.0 is output at 1V when the Amplitude is 100%.
When playback is from a Capture buffer, the input range corresponds to full scale. For
example, if a 0.1 Vpk (-20 dBV) signal is captured on an Input Range of -20 dBV (full
scale at the input), it will play back at 1V when the Amplitude is 100%. To duplicate the
actual signal amplitude, set the Amplitude to 10%.
Command: AAMP (?) {i}
Source Play Rate
Select the Arbitrary Source Play Rate. The Play Rate can be 1, 1/2, 1/4, 1/8, ... times the
maximum sampling rate. The maximum sampling rate is 262.1 kHz when the FFT Base
Frequency is 102.4 kHz and 256 kHz when the FFT Base Frequency is 100.0 kHz OR the
Measurement Group is Octave.