Display Setup Menu 4-25
SR785 Dynamic Signal Analyzer
Display Setup Menu
The Display Setup menu allows the user to choose the Measurement, View and Display
Scaling for the active display.
Measurement Group
Select the Measurement Group for both displays [FFT, Correlation, Octave Analysis,
Swept Sine, Order, Time/Histogram]. To choose a specific measurement, choose the
Measurement Group, then select a Measurement.
The FFT Measurement Group uses the Fast Fourier Transform (FFT) to compute the
frequency spectrum of finite time records.
The Correlation Measurement Group calculates the time domain auto and cross
correlation of signals.
The Octave Analysis Measurement Group measures the power within logarithmically
spaced frequency bands covering up to 11 octaves.
The Swept Sine Measurement Group uses the sine source to sweep a frequency range
making optimized measurements at discrete frequency points along the way.
The Order Measurement Group makes spectrum and time measurements which are
synchronized to the rotation of a rotating machine.
The Time/Histogram Measurement Group calculates histograms and statistical
parameters of the input signal.