Things To Watch Out For
If the analyzer is on but doesn't seem to be taking data, there are a number of things to
check.
Start
Press the [Start/Reset] key to start the measurements. Make sure the Run/Pause indicator
at the top of the screen displays ‘RUN’ instead of ‘PAUSE’. Note that in many cases
when settings are changed using the menus, the new settings will not take effect until
[Start/Reset] is pressed.
Live Display
If the displays are showing recalled trace data, they are Off-Line and do not display the
live measurement data. Set the Display to Live instead of Off-Line (in the [Display
Options] menu).
Narrow Span
If the FFT span is very narrow, the time record is very long (up to 1000’s of seconds).
Completely new data is available only every time record. Change the Time Record
Increment in the [Average] menu) to display overlapped data more often.
Low Detection Frequency
Swept Sine measurements at very low frequencies (<< 1 Hz) take a very long time (at
least 2 cycles and maybe longer). Do not set the sweep Start to a very low frequency to
measure the DC response.
Octave measurements with a very low starting band take a long time to settle before the
first measurement is valid. The settling time is related to the bandwidth of the lowest
octave band. If the Lowest Band is less than 1 Hz, the settling time can be very long.
Averaging
Very long averaging times for any measurement may give the appearance that the display
does not update. Check the FFT Number Of Averages, the Octave Integration Time or
the Swept Sine Integration Time.
When Linear averaging is on, the measurement is paused after the average is completed
(unless triggered or waterfall storage is on). Press [Start/Reset] to take another average.
Triggering
If the analyzer is waiting for a trigger, the Trig Wait indicator at the top of the screen is
on.
If the measurement is not meant to be triggered, make sure the Trigger Mode is Auto
Arm and the Trigger Source continuous.
If the measurement is meant to be triggered, make sure that the correct Trigger Source is
selected and the Trigger Level is appropriate for the trigger signal.