Real Time Bandwidth and Overlap 2-29
SR785 Dynamic Signal Analyzer
measurement is not settled. Once the 4th measurement is complete, the display returns to
full intensity since the measurements have completely settled time records.
If averaging is on, changes which unsettle the measurement will restart the average.
Unsettled measurements are not included in the new average. Averaging does not start
until the measurement is settled. When the time record increment is less than 100%,
‘Settle’ is displayed (instead of the number of averages) below the graph to indicate that
the unsettled measurements are not being averaged.
Vector Averaging
Be sure to select 100% Time Record Increment if you are interested in vector averaged
measurement results. Vector averaged results will be incorrect if anything besides 100%
Time Record Increment is used because the phase of each time record will not be the
same.
Triggering
If the measurement is triggered, then Time Record Increment is ignored. Time records
always start with the trigger (with the specified Trigger Delays). The analyzer must use
the Continuous Trigger Source to use overlap processing.