5-68 Average
Commands
SR785 Dynamic Signal Analyzer
Average Commands (FFT, Correlation, and
Order)
FAVG (?) d {, i}
The FAVG command sets (queries) the Compute Average Off/On for display d.
The parameter i selects Off (0) or On (1).
This command not valid only when the Measurement Group is Octave, Swept
Sine, or Time/Histogram. The set command requires display d to be Live.
FAVM (?) d {, i}
The FAVM command sets (queries) the type of averaging shown on display d.
The parameter i selects None (0), Vector (1), RMS (2) or Peak Hold (3).
This command is valid only when the Measurement Group is FFT, Correlation or
Order. The set command requires display d to be Live.
FAVT (?) d {, i}
The FAVT command sets (queries) the FFT Averaging Type for display d. The
parameter i selects Linear/Fixed Length (0), or Exponential/Continuous (1).
This command is valid only when the Measurement Group is FFT, Correlation,
or Order. The set command requires display d to be Live.
FAVN (?) d {, i}
The FAVN command sets (queries) the Number of Averages for display d. The
parameter i is a number of averages from 2 to 32767.
This command is valid only when the Measurement Group is FFT, Correlation,
or Order. The set command requires display d to be Live.
NAVG ? d
The NAVG command queries the Number of Averages completed for display d.
For linear averaging, the returned value is less than or equal to the FFT Number
of Averages.
For exponential averaging, the returned value is the actual number of averages
completed and eventually exceeds the FFT Number of Averages.
This command is valid in the FFT, Correlation, Order, and Octave Measurement
Groups.
FOVL (?) d {, x}
The FOVL command sets (queries) the FFT Time Record Increment for display
d. The parameter x is a percentage up to 300.
This command is valid only when the Measurement Group is FFT. The set
command requires display d to be Live.