Screen
Display
3-13
SR785 Dynamic Signal Analyzer
Octave measurements show unsettled octave bins in half intensity. Each octave bin is
displayed in half intensity until it is settled. The settling time of each bin is related to the
1/(center frequency) and the lowest bins take the longest time to settle.
Vertical Scale Bar
Chapter 3 -5 Vertical Scale Bar
The Vertical Scale Bar is at the left of the data graph for each display.
The display name, either ‘A’ or ‘B’ is at the top left corner. The display type (Live,
Done, Off-Line or Preview) is next to the display name. ‘Live’ indicates that the display
is showing live measurement results from either the inputs or from capture. 'Done'
indicates that the display is showing a completed measurement. An ‘Off-Line’ display is
showing a ‘frozen’ measurement. A display is Off-Line whenever data is recalled to the
display from a Trace or from disk. ‘Preview’ only occurs when Average Preview is On
and indicates that the display is showing a preview time record and is waiting for accept
or reject.
The Y Max and Y Min reference values are shown along the right edge. Y Max is the top
reference and Y Min is the bottom. The vertical scale per division is shown between Y
Max and Y Min. The units and scale/division are selected in the [Display Setup] menu.
The No Anti-Aliasing Filter warning is shown if any measurement input has its anti-
aliasing filter off. Measurements made without the anti-aliasing filter may contain alias
signals with frequencies above 102.4 kHz.
The Limit Test Result, either ‘Pass’ or ‘Fail’ is displayed if limit testing is on for this
display.
The Waterfall Storage Count shows how many measurement records are currently stored
in the waterfall buffer. This count is not displayed when Waterfall Storage is Off.
Y/Div Vertical Scale
Y Min Bottom
Y Max Top Reference
Display Label and Type
No Anti-Alias Filter Warning
Limit Test Result
Waterfall Storage Count