Input
Connections
2-69
SR785 Dynamic Signal Analyzer
signal behaves like a 3 µV differential signal! The CMRR decreases by about 6
dB/octave (20 dB/decade) starting at around 1 kHz.
AC vs DC Coupling
The signal input can be either AC or DC coupled. The AC coupling high pass filter
passes signals above 160 mHz (0.16 Hz) and attenuates signals at lower frequencies. AC
coupling should be used at frequencies above 160 mHz whenever possible. At lower
frequencies, DC coupling is required.
For small signals it is common that the DC offset of the signal exceeds the amplitude of
the AC signal of interest. In this case, the input range may be limited to accommodate the
large DC offset while sacrificing signal to noise in the measurement of the AC signal. If
the signal frequency exceeds 0.16 Hz, use AC coupling if possible.